Analysis of the End Point Errors Found in Mixed Acid Titrationby Conductivity Method
Abstract
On the basis of proton balance, the end point error of strong and weak acid mixture systems and its influencing factors were discussed from data of conductivity method titration. At the same time, corresponding improvements have been made in the method of end point determination. The results show that: E
t1is greatly influenced by titrant and the concentration of strong acid; E
t2is greatly influenced by the concentration of measured object and dissociation constant of the weak acid. The method of drawing a line halftime of the process is more precise than the method of drawing a line throughout the process.